Fluke 712B RTD Calibrators
Key Features
- Highly accurate, full-featured , easy-to-use single-function RTD temperature calibrator
- Measures and simulates 13 different RTD types and resistance
- Measures 4 to 20 mA signals while simultaneously sourcing a temperature signal
- Comes with a magnetic hanging tool designed in for convenient hands-free operation
- Features configurable 0% and 100% source settings for quick 25% linearity checks
- Dual inputs and backlit display for easy interpretation of measurements
- Remembers power-down settings to easily restart tests when powered up
- Provides linear ramp and 25% step auto ramp based on 0% and 100% settings
- Provides one-year and two-year specifications and traceable certification of calibration
Product overview: Fluke 712B RTD Calibrators
Fluke 712B RTD Calibrator provides highly accurate, easy-to-use, single-function RTD temperature calibration
Fluke 712B RTD Calibrator is a handheld, battery-operated instrument that measures and sources a variety of RTD types and resistances. It also includes an isolated channel to measure 4-20 mA while sourcing a temperature signal. It offers configurable 0% and 100% source settings for quick 25% linearity checks. It also provides linear ramp and 25% step auto ramp functionality based on 0% and 100% settings.
Add to that dual inputs, an easy-to-view backlit display, plus the ability to memorize settings at power down for easy restart make the 712B the ideal test tool for temperature calibration professionals who want a highly accurate, easy-to-use, single-function RTD temperature calibrator.
Other useful features:
Specifications: Fluke 712B RTD Calibrators
General Specifications | ||||||
Maximum voltage applied between any terminal and earth ground or between any two terminals | 30 V | |||||
Operating temperature | -10°C to 50°C | |||||
Storage temperature | -30°C to 60°C | |||||
Operating altitude | 2,000 m | |||||
Storage altitude | 12,000 m | |||||
Relative humidity (% RH operating without condensation) | Noncondensing | |||||
90% (10°C to 30°C) | ||||||
75% (30°C to 40°C) | ||||||
45% (40°C to 50°C) | ||||||
(Without condensation) | ||||||
Vibration requirements | MIL-T-28800E, Class 2 | |||||
Drop text requirements | 1 m | |||||
IP Rating | IEC 60529: IP52 | |||||
Electromagnetic environment | IEC 61326-1, Portable | |||||
Safety | IEC 61010-1, Max 30 V to earth, Pollution Degree 2 | |||||
Power supply | 4 AA NEDA 1.5A IEC LR6 batteries | |||||
Size (H x W x L) | 52.5 x 84 x 188.5 mm | |||||
Weight | 515 g | |||||
DC mA Measurement | ||||||
Resolution | 0-24 mA | |||||
Range | 0.001 mA | |||||
Accuracy (% of reading + counts) | 0.010% + 2 μA | |||||
Temperature coefficient | ± (0.002% of reading + 0.002% of range) /°C ( < 18°C or >28°C) | |||||
Ohms Measurement | ||||||
Ohms range | Accuracy (% of reading +counts) | |||||
0.00 Ω to 400.00 Ω | 0.015% + 0.05 Ω | |||||
400.0 Ω to 4000.0 Ω | 0.015% + 0.5 Ω | |||||
Read accuracy is based on 4-wire input. For 3-wire ohm measurements, assuming all three leads are matched, add 0.05 Ω (0.00 Ω~400.00 Ω), 0.2 Ω (400.0 Ω~4000.0 Ω) to the specifications. | ||||||
Temperature coefficient | ±(0.002% of reading + 0.002% of range) / °C (< 18°C or > 28°C) | |||||
Ohms Source | ||||||
Ohms range | 1.0 Ω to 400.0 Ω | |||||
1.00 Ω to 400.00 Ω | ||||||
400.0 Ω to 1500.0 Ω | ||||||
1500.0 Ω to 4000.0 Ω | ||||||
Excitation current from measurement device | 0.1 mA to 0.5 mA | |||||
0.5 mA to 3 mA | ||||||
0.05 mA to 0.8 mA | ||||||
0.05 mA to 0.4 mA | ||||||
Accuracy (% of reading + counts) | 0.015% + 0.1 Ω | |||||
0.015% + 0.05 Ω | ||||||
0.015% + 0.5 Ω | ||||||
0.015% + 0.5 Ω | ||||||
Resolution | 0.00 Ω to 400.00 Ω | 0.01 Ω | ||||
400.0 Ω to 4000.0 Ω | 0.1 Ω | |||||
Temperature coefficient | ± (0.002% of reading + 0.002% of range) /°C ( <18°C or >28°C) Supports pulsed transmitters and PLCs with pulse times as short as 5 ms | |||||
RTD Input and Output | ||||||
RTD Type (α) | Range (°C) | Measure (°C) | Source (°C) | |||
1-year | 2-year | Source current | 1-year | 2-year | ||
10 Ω Pt (385) | -200 to 100°C | 1.5°C | 3°C | 1 mA | 1.5°C | 3°C |
100 to 800°C | 1.8°C | 3.6°C | 1 mA | 1.8°C | 3.6°C | |
50 Ω Pt (385) | -200 to 100°C | 0.4°C | 0.7°C | 1 mA | 0.4°C | 0.7°C |
100 to 800°C | 0.5°C | 0.8°C | 1 mA | 0.5°C | 0.8°C | |
100 Ω Pt (385) | -200 to 100°C | 0.2°C | 0.4°C | 1 mA | 0.2°C | 0.4°C |
100 to 800°C | 0.015% +0.18°C | 0.03% +0.36°C | 1 mA | 0.015% +0.18°C | 0.03% +0.36°C | |
200 Ω Pt (385) | -200 to 100°C | 0.2°C | 0.4°C | 500 μA | 0.2°C | 0.4°C |
100 to 630°C | 0.015% +0.18°C | 0.03% +0.36°C | 500 μA | 0.015% +0.18°C | 0.03% +0.36°C | |
500 Ω Pt (385) | -200 to 100°C | 0.3°C | 0.6°C | 250 μA | 0.3°C | 0.6°C |
100 to 630°C | 0.015% +0.28°C | 0.03% +0.56°C | 250 μA | 0.015% +0.28°C | 0.03% +0.56°C | |
1000 Ω Pt (385) | -200 to 100°C | 0.2°C | 0.4°C | 250 μA | 0.2°C | 0.4°C |
100 to 630°C | 0.015% +0.18°C | 0.03% +0.36°C | 250 μA | 0.015% +0.18°C | 0.03% +0.36°C | |
100 Ω Pt (3916) | -200 to 100°C | 0.2°C | 0.4°C | 1 mA | 0.2°C | 0.4°C |
100 to 630°C | 0.015% +0.18°C | 0.03% +0.36°C | 1 mA | 0.015% +0.18°C | 0.03% +0.36°C | |
100 Ω Pt (3926) | -200 to 100°C | 0.2°C | 0.4°C | 1 mA | 0.2°C | 0.4°C |
100 to 630°C | 0.015% +0.18°C | 0.03% +0.36°C | 1 mA | 0.015% +0.18°C | 0.03% +0.36°C | |
10 Ω Cu (427) | -100 to 260°C | 1.5°C | 3°C | 1 mA | 1.5°C | 3°C |
120 Ω Ni (672) | -80 to 260°C | 0.15°C | 0.3°C | 1 mA | 0.15°C | 0.3°C |
50 Ω Cu (427) | -180 to 200°C | 0.4°C | 0.7°C | 1 mA | 0.4°C | 0.7°C |
100 Ω Cu (427) | -180 to 200°C | 0.2°C | 0.4°C | 1 mA | 0.2°C | 0.4°C |
YSI400 | 15 to 50°C | 0.2°C | 0.4°C | 250 μA | 0.2°C | 0.4°C |
1. Sensor inaccuracies not included. 2. Resolution: 0.1°C. 3. Read accuracy is based on 4-wire input. For 3-wire RTD measurements, assuming all three RTD leads are matched, add 1.0°C (Pt10 and Cu10), 0.6°C (Pt50 and Cu50), 0.4°C (Other RTD types) to the specifications. 4. Source Accuracy in source mode is based on 0.5 mA~3 mA (1.00 Ω~400.00 Ω), 0.05 mA~0.8 mA (400.0 Ω~1500.0 Ω), 0.05 mA~0.4 mA (1500.0 Ω ~4000.0 Ω), excitation current (0.25 mA for Pt1000 range). 5. Temperature Coefficient: ±0.05°C /°C for measure, ±0.05°C /°C (< 18°C or > 28°C) for source. 6. Supports pulsed transmitters and PLCs with pulse times as short as 5 ms. |
Models: Fluke 712B RTD Calibrators
Fluke 712B
Fluke 712B RTD Calibrator
Includes:
- Magnetic hanging tool
- Batteries
- Manual
- Traceable calibration certificate
- Test Leads
Manuals + resources: Fluke 712B RTD Calibrators
Manuals (9)
712B | Users Manual
712B | Users manual supplement